► AEC ▼ Rotor-AID
|
|
ROTOR-AID LC100 SPECIFICATIONS
Standard Configuration: Two-Tube radiographic and fluoroscopic 'R' stator (19" rack mount) Power Requirements: Nominal 20 Ampere service for 'R' stators and 30 Ampere service for 'Q' stators, single phase 200VAC to 240VAC +5% - 10%, 50/60Hz Power Output: Voltage output selected by transformer taps. Nominal 'R' stator voltages: High Speed Boost 450VAC Maximum available power output: For High Speed - nominally 3kW per phase into most 3-wire split stator windings rated for intermittent power levels of up to 500VAC and up to 10 Amperes per winding. The output frequency is 180Hz For Low Speed - nominally 2.5kW per phase into most 3-wire split stator windings rated for intermittent power levels of up to 250VAC and up to 10 Amperes per winding. The output frequency matches the line frequency (50/60 Hz). Timer Adjustments: Individual timers are provided for each x-ray tube to set boost and brake times. Additional timers are available to set continuance (holdover time) and to set time for other functions such as spot hold, and cassette delay. Physical Parameters: 19" Rack Mount inches - 5.25H x 19W x 12D mm - 133H x 483W x 305D Weight - 40 lbs. (18 kg) Environmental Characteristics Ambient Temperature, storage: -20 to +65°
C
Certification: All units ETL certified. Units may also be CETL and IEC601.1 certified when requested and so labeled. Also available for special applications: Custom enclosures |
||||