Advanced Instrument Development, Inc. 2545 CURTISS STREET DOWNERS GROVE, ILLINOIS 60515 USA TEL: 630-271-9729 FAX: 630-271-9995
|
|||||
|
► AEC ▼ Rotor-AID
|
|
ROTOR-AID RC100 SPECIFICATIONS
Standard Configuration: Single tube radiographic (19-inch rack mount) Three-tube radiographic and fluoroscopic (19-inch rack mount) Options: Substitution of input isolation transformer (adds 5.25"/133mm to
height of rack mount; adds 8 lbs./ 4 kg. to
overall weight) Substitution of 277VAC line filter in place of standard 240VAC line
filter Decorative wall mount cabinet Power Requirements: Nominal 30 Amperes, 200-240 +/-10% VAC, 50/60 Hz service required
(277VAC with optional line filter). Unit includes 20 Amperes, 240 VAC circuit breaker and 240V line
filter. Power Output: Fixed voltage output to tube stator with programmable duty cycles.
Changing the duty cycles of the output waveform provides proper stator
currents during boost, run and brake. Maximum available power output: Nominally 3kW per phase into most 3 wire split-phase stators rated
for intermittent power levels of up to 550 VAC and up to 10 Amperes per
winding. Drive frequencies for boost and run are selectable in 2.5 Hz
increments; Low Speed: 25-75.2 Hz High Speed: 100-197.5 Hz Timer Adjustments: Individual switches are provided for each x-ray tube to set boost and
brake times. Additional switches are provided to set continuance
(holdover time) and to set time for an auxiliary function. Physical Parameters: 19" Rack Mount inches - 8.75H x 19W x 11.88D mm - 222H x 483W x 302D Weight - 55 lbs. (25 kg) Optional Rack Mount Cabinet inches - 19H x 20W x 15.5D mm - 483H x 508W x 394D Certification: CETL and IEC601.1 Certified, when requested and labeled Also available for special applications: Rotor controllers for three-phase x-ray tube stators. Custom enclosures |
||||
Copyright 2012, Advanced Instrument Development, Inc.